To check the output of the
DC CALIBRATOR
In this test you measure the output of the
DC CALIBRATOR
with a multimeter.
The
DC CALIBRATOR
is used for self-calibration of the oscilloscope. The
accuracy is not specified, but it must be within the test limits to provide for
accurate self-calibration.
Test limits
5.000 V
±
10 mV and 0.000 V
±
500
µ
V.
Table 4-2
Equipment Required
Equipment Critical specifications Recommended
Model/Part
Digital Multimeter 0.1 mV resolution, better than 0.01% accuracy HP 34401A
Cable BNC HP 10503A
1
Connect a multimeter to the rear panel
DC CALIBRATOR
connector.
2
Press
Print/Utility
.
3
Press the
Self Test
softkey, then press the
DAC
softkey.
The multimeter should measure 0.00 V dc
±
500
µ
V. If the result is not
within the test limits, see "Troubleshooting the oscilloscope," on page 4–40.
4
Press any key to continue the test.
The multimeter should read 5.000 V
±
10 mV. If the result is not within the
test limits, see "Troubleshooting the oscilloscope," on page 4–40.
Service
Verifying Oscilloscope Performance
4–6
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...