To verify Vertical Output on Option 005
This section applies only to Option 005 Enhanced TV/Video Trigger if installed on
the HP 54602B.
In this test we will use the oscilloscope’s channel 2 to measure the amplitude
of the Vertical Output (
VERT OUT
connector on rear panel) signal.
Test limits: ~90 mVp-p into 50
Ω
with a full screen input.
Table 4-8
Equipment Required
Equipment
Critical specifications
Recommended Model/Part
Signal generator
1 to 500 MHz at 200 mV
HP 8656B opt 001
Cable
BNC, 48 inch
HP 10503A
Cable
Type N (m), 24 inch
HP 11500B
Adapter
Type N (m) to BNC (f)
HP 1251-0780
1
Connect the signal generator to oscilloscope channel 1 input.
2
Set the signal generator to equal the full bandwidth of your
oscilloscope, and set the output level to 0 dBm.
3
Connect the signal generator to oscilloscope channel 1. Set channel
1
Input
to
50
Ω
to correctly terminate the signal generator.
4
Press
Autoscale
.
5
Adjust the oscilloscope controls and signal generator to obtain an
8-division high display.
6
Connect oscilloscope rear-panel
VERT OUT
to oscilloscope channel 2.
Set channel 2
Input
to
50
Ω
mode.
7
Measure the peak-to-peak amplitude of channel 2. It should be
greater than or equal to 63.6 mVp-p.
Because the measurement is being made at the full bandwidth of the
oscilloscope’s channel, the peak-to-peak measurement is corrected for the
oscilloscope’s high frequency roll off.
Service
Verifying Oscilloscope Performance
4–22
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...