1
Set up the oscilloscope.
a
Press
Setup
, then press the
Default Setup
softkey.
b
Press
Voltage
, then press the
V avg
softkey.
c
Set the Volts/Div to the first line of table 4-4.
d
Adjust the channel 1 Position knob to place the baseline near
(but not at) the bottom of the display.
2
Press
Cursors
, then press the
V1
softkey.
3
Using the cursors knob, set the V1 cursor on the baseline.
If you are in an electrically noisy environment, it can help to place a shorting
cap on the input BNC connector when positioning V1.
4
Connect the power supply to the oscilloscope and to the multimeter,
using the BNC tee and cables.
5
Set the power supply output to the first line in table 4-4.
Service
Verifying Oscilloscope Performance
4–8
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...