To adjust the display
When to adjust the display
The display adjustments are optional and normally do not require adjustment.
You should use this procedure only for the few cases when the display is
obviously out of adjustment.
Table 4-16
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Digital multimeter
Accuracy
±
0.05%, 1 mV resolution
HP 34401A
1
Turn off the oscilloscope and disconnect the power cord.
2
Remove the cover from the oscilloscope.
3
Reconnect the power cord and turn on the oscilloscope.
4
Connect the digital multimeter to the end of R901 closest to the fuse.
See figure on next page.
5
Adjust
+
B for
+
14.00 V.
6
Press
Print/Utility
. Press the
Self Test
Softkey, then press
the
Display
softkey.
7
Adjust V.HO (vertical hold) for vertical synchronization.
8
Set the intensity control (on the front panel) to mid-range.
9
Adjust Sub Bri (sub bright) to the lowest setting so that the half
bright blocks on the display are visible.
10
Increase the intensity control to a comfortable viewing level.
This is usually about 3/4 of its maximum range.
Service
Adjusting the Oscilloscope
4–37
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...