Vertical self cal
3
After the message
"Default calibration factors loaded"
is
displayed on the lower left side of the display, press the
Vertical
softkey.
4
Follow the instructions on the display, then press the
Continue
softkey.
The display prompts instruct you to connect the rear panel
DC
CALIBRATOR
output first to channel 3, then to channel 1, then to channel 4, and finally to
channel 2. (Channels 1 and 2 only on the HP 54600B and HP 54603B.)
5
After the message
"Press Continue to return to calibration
menu"
appears on the display, press the
Continue
softkey.
6
Connect a pulse generator set to 100 kHz and 1 V p-p and with a rise
time less than 5 ns to channels 1 and 2. Place the feedthrough at the
BNC tee. Make sure you use the HP 10502A cables to ensure equal
cable lengths.
Delay self ca
l
7
Press the
Delay
softkey, then follow the instructions on the display.
The display will instruct you to connect the signal simultaneously to channels
1 and 2, 1 and 3, 1 and 4, 2 and 4, 2 and 3, and finally to 3 and 4.
8
Set the rear-panel
CALIBRATION
switch to
PROTECTED
.
Service
Adjusting the Oscilloscope
4–32
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...