To reset the instrument setup
1
To reset the instrument to the default factory-preset configuration,
press
Setup
.
2
Press the
Default Setup
softkey.
3
To reset the instrument to the configuration that was present before
pressing
Autoscale
, press the
Undo Autoscale
softkey.
Table 2-1
Default Setup configuration settings
Configuration Item
Setting
Cursors
Cursors off; time readout is selected; all cursors are set to time/voltage zero.
Trace memories
Both trace memory 1 and 2 are off; trace 1 memory is selected.
Setup memories
Setup memories are off; setup memory 1 is selected.
Graticule
Grid set to Full
Autostore
Off
Time base
Time reference center; main, not delayed sweep; main and delay value 0; 100
µ
s/div
main time base.
Display
Vectors On, Display Mode Normal.
Channels
Channel 1 on, Position 0 V, Volts/Div 100 mV.
Trigger Mode
Auto Level, Coupling DC, Reject Off, Noise Reject Off.
Trigger Condition
Rising edge of channel 1
Operating Your Oscilloscope
To reset the instrument setup
2–33
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...