To check the Low Voltage Power Supply
1
Disconnect the power cord, then set the oscilloscope on its side.
2
Connect the negative lead of the multimeter to a ground point on the
oscilloscope. Connect the power cord and turn on the oscilloscope.
3
Measure the power supply voltages at L
1
, L
2
, and L
3
on the system
board.
Table 4-19
Power Supply Voltage Tolerances
Supply Voltage Tolerance
+
5.1 V
±
150 mV (
+
4.95 V to
+
5.25 V)
+
15.75 V
±
787 mV (
+
14.96 V to
+
16.54 V)
−
15.75 V
±
787 mV (
−
14.96 V to
−
16.54 V)
Low Voltage Power Supply voltage test points
If the
+
5.1 V measurement is out of the test limits, adjust the
+
5.1 V
adjustment on the power supply. The
±
15 V supplies are not adjustable and
are dependent upon the
+
5.1 V supply.
Blown fuse
If the fuse is blown in the power supply, the power supply is defective. Replace
the power supply.
Figure 4–7
Service
Troubleshooting the Oscilloscope
4–45
Содержание HP 54600B
Страница 7: ...vi ...
Страница 12: ...1 The Oscilloscope at a Glance ...
Страница 16: ...Softkey Menu Reference Press this key To obtain this menu Figure 1 4 To obtain this menu Press this key 1 5 ...
Страница 28: ...2 Operating Your Oscilloscope ...
Страница 69: ...2 42 ...
Страница 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Страница 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Страница 151: ...Exploded view of Option 005 and related oscilloscope parts Figure 4 12 Service Replacing Parts in the Oscilloscope 4 62 ...
Страница 153: ...4 64 ...
Страница 154: ...5 Performance Characteristics ...
Страница 175: ...Glossary 8 ...