Chapter 6. Continuous
82
Hospira GemStar System Operating Manual
Preparing for Secondary Delivery (Piggybacking)
1
Ensure a primary piggyback pump set with a
proximal Y-site backcheck valve is in place.
2
Use the extension hook from the secondary set
to suspend the primary container from the I.V.
pole.
3
Prepare the secondary set according to the
instructions included with the set.
4
Attach a needle or blunt cannula (if appropriate)
and prime the secondary set.
5
Suspend the secondary container from the I.V.
pole.
6
Connect the secondary set to the proximal
Y-site (or proximal luer port).
d
Note: When using a secondary container of
500 mL or greater, ensure the bottom of the
secondary container is at least 7 inches (17.8
cm) above the fluid level in the primary
container. Use additional extension hooks if
necessary.
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...