Chapter 17. GemStar Pump Specifications
188
Hospira GemStar System Operating Manual
Occlusion Information
Avoiding Bolus Infusion After Occlusion
1
Press [
STOP
].
2
Close the distal slide clamp or lower CAIR
(roller) clamp.
3
Press the cassette release button to eject the
cassette from the pump.
4
Press down on
to open the flow stop.
5
Wait 10 seconds.
6
Press down on
to close the flow stop.
7
Ensure slide clamp or lower CAIR clamp is
open prior to inserting cassette.
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...