Hospira GemStar System Operating Manual
55
Chapter 3. Total Parenteral Nutrition
Changing a TPN Program
To change a program:
1
From the
STOP
mode, press [
CHANGE
] and select
CHANGE PROGRAM
.
2
As each message displays, make the desired changes or press [
YES
/
ENTER
] to
accept the current setting.
3
Review the program. Press
to begin review.
4
Press [
START
] to begin infusion.
Can be changed before
is pressed
Can be changed at any
time
Cannot be
changed
•
VTBI
•
Taper times
•
Total TPN time
•
KVO rate
•
Increase container size
only
•
Air sensitivity
•
Delivery mode
d
Note: You cannot change a program when the pump is in
FULL
or
CONTAINER
Lock.
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...