Chapter 12. Troubleshooting
150
Hospira GemStar System Operating Manual
PROGRAMMING
INCOMPLETE
N
A/V
Programming has not
been completed and no
keys have been pressed
within three minutes.
1
Press any key to clear
the alarm.
2
Complete the program
entries.
START
•
START flashes on
the display. Other
parts of the display
are not affected.
N
A/V
A Stop Mode Callback
has occurred because the
programmed pump has
remained in the STOP
mode for more than three
minutes.
1
Press [
SILENCE
] to
quiet the alarm for 3
minutes.
2
Follow the instructions
on the display to clear
the alarm.
Alert Display Message
Is the
pump
infusing?
Alarm Cause/Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...