Chapter 2. Basic Operation
38
Hospira GemStar System Operating Manual
7
Push down on the flow stop
to
close the cassette.
8
Ensure slide clamp is open prior to
inserting cassette.
9
Load the cassette.
CAUTION: Ensure all four
cassette latches are clearly
visible after the cassette is
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed, DO NOT use
the pump set.
10
Connect the pump set to the
patient-access device.
Priming using [
PURGE
] is only
allowed before [
START
] is pressed
or after one of the following
events occur when using Pain
Management therapy:
6
Complete Review of New
Program
6
New Container
6
Air-in-Line Alarm
d
Note: Ensure the tubing is behind
the tubing guide (at the distal
end).
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...