Chapter 2. Basic Operation
52
Hospira GemStar System Operating Manual
12
Press [
PURGE
].
13
The pump displays
PRIME THE
SET?
Press [
YES/ENTER
].
14
Press and hold [
PURGE
].
Continue using the [
PURGE
] key
until fluid fills the tubing. Ensure
all air is removed from the
container, cassette, and tubing.
CAUTION: When priming is
complete, ensure no fluid flows
at the distal end of the pump set.
If flow is observed,
DO NOT use the pump set.
15
Connect the pump set to the
patient-access device.
d
Note: Purge rate is 250 mL/hr.
The pump purges for up to 2
minutes each time you press and
hold
[
PURGE
]
.
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...