Appendix 1 Error Messages
A
7
Appe
ndix
512
A locking violation occurred while ac-
cessing (file name).
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
513
Disk full while accessing (file name).
Saving is not possible because of insuffi-
cient space on the storage media. Delete
files or replace the storage media. If mea-
suring, stop measurement, then replace
the storage media.
"11.7.3 Deleting Files &
Folders" (p. 307)
514
An attempt was made to access (file
name) past its end.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
520
(File name) may be corrupted.
The file may be corrupted. This file cannot
be used.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
Displayed Errors
Msg
No.
Message
Remedial Action
93
Measurement is not possible with the
current module configuration. Install
no more than four 8961s.
Up to four 8961 High Voltage Units (modules) may be used. Install
no more than four 8961s. (8861-50)
160
Measurement was aborted.
(Cause) Real-time measurement and saving operations were
aborted due to an error. Remove the cause of the error.
170
Unsupported FPGA version detected.
Upgrade to a compatible FPGA.
Upgrade to an FPGA version that is supported by the application
program.
171
Unsupported Kernel version detected.
Upgrade to a compatible Kernel.
Upgrade to a kernel version that is supported by the application
program.
176
Internal temperature is abnormal.
Please turn the power off after preserv-
ing necessary data.
Internal temperature is abnormal. Measurement has been
stopped. Please turn the power off after preserving necessary da-
ta. If there is no problem with the use temperature and installation
environments, then there is a possibility that the instrument may
be damaged. Please send it for repair.
177
Input unit temperature is rising. Please
confirm the use temperature and the
installation environment.
Internal temperature is abnormal. Please turn the power off after
preserving necessary data. If there is no problem with the use
temperature and installation environments, then there is a possi-
bility that the instrument may be damaged. Please send it for re-
pair.
180
USB host controller malfunction de-
tected.
Have the instrument repaired.
181
Keyboard controller malfunction de-
tected.
Have the instrument repaired.
182
I/O FPGA malfunction detected.
Have the instrument repaired.
183
Storage FPGA malfunction detected.
Have the instrument repaired.
184
Bus bridge malfunction detected.
Have the instrument repaired.
190
Module power supply malfunction de-
tected.
Power to the modules was momentarily interrupted. If this oc-
curred while measuring, data may have been corrupted.
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......