9.2 Setting and Analysis Workflow
238
Measurement
Configuration Settings
Make settings on the Status Settings screen.
Measurement using different sampling rates is not available
with the Real-Time Saving function.
Press the
SET
key to open the Settings screen
Press the
SUB MENU
keys to select the
menu
Press the
SHEET/PAGE
keys
to select the
[Use Ch]
page
1
Selecting channels to use
With the Real-Time Saving function, measure-
ment using different sampling rates simulta-
neously is not available. Also, the Model 8958
16-Ch Scanner Unit cannot be used.
Press the
SHEET/PAGE
keys
to select the
[Basic]
page
2
Making settings for saving data
Set the save destination and file name.
Verify that there is adequate space available at
the save destination, and that the media is in-
stalled or inserted properly.
3
Make measurement waveform settings
Select the timebase (sampling rate) and re-
cording length (duration).
The data acquisition interval on the time axis is
set by the timebase or sampling rate setting.
Recording length can be set as a number of di-
visions or recording duration.
4
Make whole waveform settings
The timebase can be set automatically or man-
ually.
5
Set the trigger mode (recording method)
Select single (one-shot), continuous or timer.
1
2
4
5
3
You can verify the free space on the save destination and the file
size to be saved using the current settings. Also, the maximum re-
cording length can be set to match the free space on the save des-
tination storage media.
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......