11.3 Saving Data
272
11.3 Saving Data
11.3.1 Save Sequence
After installation and connection,
turn the power on.
Measurement Preparations
See
"Chapter 3 Measurement Preparations" in the
Quick Start Manual
Set the measurement configuration.
Set the measurement configuration on each settings screen ([Status],
[Channel], [Trigger] menu).
Insert the storage media.
Before inserting the storage media, make sure write protection is dis-
abled.
See
(Initialize the storage media.)
When using unformatted storage media, initialize the storage media
from the File screen.
See
"11.1.5 Initializing (Formatting) Storage Media" (p. 266)
Select the save method.
Configure the settings on the Save Settings screen ([Save] menu).
See
"11.3.2 Save Methods" (p. 273)
Select whether to save data automatically during measurement or set the
data to save and perform manual saving (SAVE key) after measurement.
• When saving automatically:
[Auto Save]
page (p. 276)
• When saving manually: [
SAVE key]
page (p. 278)
With
[Selection Save]
, the data is saved after selecting the save destina-
tion and the data to save in the dialog box that appears upon pressing
the
SAVE
key.
With
[Quick Save]
, the data is saved upon pressing the
SAVE
key if the
data to save is preset in the Settings screen.
Settings data can be saved regardless of whether measurement has not
begun or has ended.
Select the save destination.
Make sure the storage media has been inserted in the instrument.
Select the storage media and save destination in the dialog box.
(p. 275)
Select the data to save.
Set the data to save such as waveforms, numerical calculations, and
screens.
The data that can be saved differs depending on whether automatic saving
or manual saving is performed.
See
"Data Capable of Being Saved" (p. 273))
(*: Manual saving only)
For auto save, make sure the auto save setting is set to On before be-
ginning measurement.
Auto Save
The data is saved before and after measurement.
To stop saving, press the
STOP
key. Measurement also ends at the same
time.
Manual Save
Quick Save:The data is saved upon pressing the
SAVE
key.
Selection Save:The data is saved after setting the data to save and per-
forming the save operation in the dialog box that appears upon pressing the
SAVE
key.
Manual Save
Selection
Save
Quick Save
(Measure)
SAVE key
Auto Save
Auto Save
Manual Save
• Waveform data
• Settings data*
• Numerical calcu-
lation results
• Display screen*
Measure
SAVE key
Execute button
(Measure)
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......