Appendix 1 Error Messages
A
4
67
Too many measurement channels.
You have tried to use more channels than
the number enabled for use. Either in-
crease the number of channels enabled
for use, or turn unused channels Off.
"4.2.1 Selecting Chan-
nels to Use" (p. 86)
68
The offset value is too large for Offset
Cancel.
Offset Cancel is not available if the input
signal is more than ±10 divisions from 0 V.
Change the range, and execute Channel
Offset again.
Input Module Guide
:
"3.11.18 Executing Off-
set Cancellation"
69
Up to six Model 9322 Differential Probes
can be used (with Model 9325)
When the Model 9322 Differential Probe is
connected with the Model 9325 Power
Cord, up to six channels can be used si-
multaneously.
Do not use more Model 9322 Differential
Probes than this limitation allows.
70
Voltage Sag triggering is disabled. (Val-
id time base range: 20 µs/div to 50 ms/
div)
Voltage Sag triggering can be used only
when the time base is between 20 µs/div
and 50 ms/div.
71
Voltage Sag triggering is disabled for
scanner modules.
Voltage Sag triggering cannot be used
with a scanner module.
72
Zero-adjustment failed.
Execute zero-adjustment again.
73
Offset Cancel failed.
The input voltage is more than ±10 divi-
sions from 0 V. Set the input signal within
±10 divisions from 0 V.
Input Module Guide
:
"3.11.18 Executing Offset
Cancellation"
74
Auto balance failed.
Check whether a sensor is in an un-
charged state, and that it is connected cor-
rectly.
Input Module Guide
:
"3.11.19 Executing Auto-
Balance"
75
Time base can be set from 20 ms/div.
When the Recording Length is set to
[Cont] with the Recorder function, the time
base must be at least 20 ms/div.
76
Measurement is not possible with the
current module configuration. (REC
Function)
(Model 8861-50 only)
When four Model 8946 4-Ch Analog Units
are installed in Unit locations 1 to 4 and a
Model 8958 16-Ch Scanner Unit is also in-
stalled, the Recorder function cannot be
used for measurement. Install no more
than three Model 8946s in Unit locations 1
to 4.
78
No waveform, or recording length is too
long.
If no waveform is present, execute mea-
surement.
If the recording length is too long, perform
a partial save, then reload and calculate.
79
Measurement is not possible with the
current module configuration. (FFT
function)
FFT function measurements are not avail-
able when only the Model 8958 16-Ch
Scanner Unit is installed.
80
The time base and sampling rate cannot
be changed during synchronized mea-
surement.
During synchronized measurement sam-
pling, the time base and sampling rate
cannot be changed. Finish measuring,
change the time base or sampling rate,
and resume synchronized measurement
sampling.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......