4.2 Setting Measurement Configuration (Status Settings Screen)
97
4
Chapte
r 4
Me
asuremen
t Co
nfiguration Sett
ings
Set the length (number of divisions) to record each time data is acquired.
The following methods and settings are available:
•
Fixed recording length [Fixed]:
select from the fixed recording lengths
•
Set arbitrary recording length [User]:
set an arbitrary recording length in
units of divisions (p. 99).
•
Continuous [Cont]:
records continuously (Recorder Function only) (p. 101).
Recording Length and Data Samples
• Memory Function and memory waveforms with the REC&MEM function.
Each division of the recording length consists of 100 data samples. The total
number of data samples for a specified recording length = set recording length
(divisions) × 100 + 1.
• Recorder Function and recorder waveforms with the REC&MEM function.
Each recording length division = 100 pairs of data points, with each pair com-
posed of two samples: the maximum and minimum measured values within
each sampling period.
See
"Appendix 4.4 Recorder Function Values" (p. A50)
4.2.4
Setting the Recording Length (number of divi-
sions)
To change recording length while measuring
Recording length can be changed on the Waveform or Settings screens. The re-
cording length becomes effective at the time the setting is changed.
See
Modifying the Waveform screen view: "4.4 Setting Measurement Configuration on
the Waveform Screen" (p. 114)
3
2
Setting a Fixed Recording Length (Fixed Shot)
To open the screen: Press the
SET
key
Select
with the
SUB MENU
keys
Status Settings screen
See
Screen Layout (p. 28), To set from the Waveform screen (p. 114)
Operating Key
Procedure
1
(with the Memory function)
SHEET/PAGE
Select the
[Basic]
page.
2
Select the setting method for recording length.
CURSOR
Move the cursor to the
[Shot]
item.
F1
Select
[Fixed]
.
3
Set the recording length.
CURSOR
Move the cursor to the
[Fixed Shot]
(Fixed re-
cording length) item.
F1 to F8
(Switch Display: F8)
Select the length of waveform to be acquired (re-
cording length).
Displayed recording time and maximum
recording length are linked to the record-
ing length setting.
1
(Memory Function case)
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......