![Hioki 8860-50 Скачать руководство пользователя страница 432](http://html1.mh-extra.com/html/hioki/8860-50/8860-50_instruction-manual_2137112432.webp)
Appendix 1 Error Messages
A
6
203
No response from server.
Verify the network settings on the PC at
the connection destination.
Analysis and
Communication
Supplement
"Chapter 4
Communications Settings"
210
Cannot connect to the mail server.
Check the mail server settings.
211
Communication error with the mail serv-
er.
Check the mail sending settings.
212
Fail to send mail.
Check the mail sending settings.
213
Mail spool over.
Check the mail sending settings.
214
POP authorization needed.
Configure POP authentification.
215
Invalid POP account or password.
Enter the correct user name and pass-
word.
216
Invalid mail settings.
Check the mail sending settings.
501
An unknown error occurred while ac-
cessing (file name) .
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
502
(File name) was not found.
Verify that the file saving destination or file
to load is correct. Alternatively, the free
space of the directory can be insufficient.
In this case, either delete files in the sav-
ing destination directory, or change to an-
other saving destination.
"When the maximum
number of files that can
be saved is exceed-
ed:"(p. 277)
503
(File name) contains an invalid path.
Verify that the file saving destination or file
to load is correct.
"11.2 Data Capable of
Being Saved & Loaded"
(p. 267)
504
(File name) could not be opened be-
cause there are too many open files.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
505
Access to (file name) was denied.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
506
An invalid file handle was associated
with (file name) .
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
507
(File name) could not be removed be-
cause it is the current directory.
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
508
(File name) could not be created be-
cause the directory is full.
Since the number of files in the saving
destination directory is exceeding the
maximum number of files, the file cannot
be saved.
Either delete files in the saving destination
directory, or change to another saving
destination.
"When the maximum
number of files that can
be saved is exceed-
ed:"(p. 277)
509
Seek failed on (file name).
An internal fault may have occurred in the
instrument. Turn the instrument off and
back on.
510
A hardware I/O error was reported while
accessing (file name).
The storage media may be corrupted. Re-
place with new storage media.
511
A sharing violation occurred while ac-
cessing (file name).
Verify the settings (user name and pass-
word) of the shared destination.
Displayed Warnings
Msg
No.
Message
Remedial Action
Reference
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......