4.3 Acquiring Waveforms Using the Utility Functions
113
4
Chapte
r 4
Me
asuremen
t Co
nfiguration Sett
ings
• The times during which sampling is inhibited (dead time) due to display and
recording processing after each block of data has been acquired are about 8
ms.
• When measuring a parameter other than voltage or current with the Model
8940 F/V Unit, dead time is about 230 ms.
• When using the Model 8958 16-Ch Scanner Unit or Timebase 2 sampling,
dead time may be longer, depending on the Timebase 2 sampling speed set-
ting.
• When the Trace Waveform display is disabled, even if the Roll Mode is
enabled (other than Off), the Roll Mode function is unusable.
• The Averaging function is not available when Memory Division is enabled.
• When triggering occurs very often, pressing the STOP key may not stop mea-
surement until enough data has been acquired to fill the blocks specified for
use.
Содержание 8860-50
Страница 2: ......
Страница 12: ...Contents x...
Страница 18: ...Reading this Manual 6...
Страница 24: ...1 3 Interconnection and Block Diagrams 12...
Страница 180: ...6 12 Making Trigger Settings on the Waveform Screen 168...
Страница 246: ...8 15 Inserting and Searching for Event Marks on a Waveform 234...
Страница 354: ...12 7 Print Examples 342 Print Example 4 Row Printing 1 4 steps Settings data Each step is 25 div...
Страница 356: ...12 7 Print Examples 344 A B Waveform_________________________________________________...
Страница 360: ...12 7 Print Examples 348 Event ________________________________________________________ Event mark...
Страница 406: ...14 2 External I O 394...
Страница 489: ......
Страница 490: ......