101 / 174
en
12.2020 / Rev. 02
AB66005
Operation
Fig. 158: Diagram of ‘Temperature guard’, class 3.3
A1 Target temperature 1
B1 Temperature specification, class 3.1
C1 Deviation from the target temperature starting from time C1
D1 Class 3.1 active starting from time D1
A2 Target temperature 2
B2 Temperature specification, class 3.2
C2 Deviation from the target temperature starting from time C2
D2 Class 3.2 active starting from time D2
Contradiction between the temperature setpoint and
temperature guard setting
Manual mode: Manual mode can be started even if
there is a contradiction between the temperature
setpoint and the guard temperature. A warning mes-
sage appears.
Program mode: After the program starts, a warning
message appears: ‘The programmed temperature is
higher than the set limit value of the temperature
guard ...’. After confirming this message, it leads you
back to the program start settings.
6.13.6
Control contact (option)
The
[Control contact] button is only available for incubators with
either the Switchboard option or the passive dehumidification option.