Cleaning, disinfection and maintenance
1
Fig. 212: Right baffle sheet
1 Right baffle sheet
7.
Disengage the right baffle sheet upward and then remove
toward the front.
Installation is done in the opposite order.
When installing the upper baffle sheet, push this upward and fasten
with the two knurled screws. It must be flush with the ceiling of the
interior.
7.2
Disinfection
■
If infectious material gets into the interior, disinfect it immedi-
ately.
■
Ingredients of suitable disinfectants: Ethanol, n-propanol,
isopropyl alcohol, ethyl hexanol, corrosion inhibitors.
■
After using disinfectants, remove the disinfectant residue by
wiping with a damp cloth.
■
The surfaces must be dried immediately after disinfecting.
WARNING
Risk due to insufficient disinfection after contamina-
tion!
There is a risk of infection for the user. Observe your
laboratory rules (hygiene plan, etc.) for handling bio-
logical agents.
7.3
Removal of radioactive contaminants
■
The agent must be specifically labelled as being an agent
for removing radioactive contaminants.
■
Ingredients of suitable agents for removing radioactive con-
taminants: anionic tensides, non-ionic tensides.
■
After removing the radioactive contaminants, remove the
agent residue by wiping with a damp cloth.
■
The surfaces must be dried directly after removing the radi-
oactive contaminants.
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...