137 / 174
en
12.2020 / Rev. 02
AB66005
Technical data
EMC
Emitted interference, immunity to
interference
EN/IEC 61326-2-6, Class B
Type of protection (EN 60529)
IP 20
Temperatures
Temperature range
1 K above the ambient temperature up to 65 °C
Temperature setting precision
0.1°C
Temperature deviation over time at
37 °C
±
0.1 K
Spatial temperature deviation at
37 °C
±
0.2 K
Spatial temperature deviation at
25 °C
±
0.1 K
Recovery time (after door open for
30 s) at temperature setpoint 37
°C
≤
4.5 min
Dimensions
Interior dimensions
W x D x H
535 x 690 x 850 mm
Exterior dimensions
W x D
1)
x H
710 x 825
1
x 1425 mm
Weight
approx. 117 kg
3)
Maximum load per standard feed
50 kg
Maximum total load
100 kg
1)
Without door handle and access port (
Ø
42 mm).
2)
With Switchboard option (external socket strip)
3)
With glass door +10 kg.
All temperature data was measured at an ambient temperature of
22°C and in acc. with DIN 12880:2007-05. The data applies to incu-
bators with standard equipment.
9.4
HettCube 400 R
Manufacturer
Andreas Hettich GmbH & Co. KG, D-78532 Tuttlingen
Model
HettCube 400 R
Type
64005
64005-01
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...