43 / 174
en
12.2020 / Rev. 02
AB66005
Operation
Before operation, complete commissioning in full
Chapter 5 ‘Commissioning’ on page 27.
After a malfunction (power failure, door open too
long, etc.), the operating company must assess
whether the samples are still usable.
6.1
Operating elements
1
3
2
4
Fig. 48: Operating elements, front side
1 Door lock
2 Service interface (USB)
3
[Incubator On/Off] switch
4 Touchscreen
6.2
Loading
The maximum load per standard slide-in module is
50 kg.
The maximum load per slide-in module or drawer
with telescopic rails is 40 kg.
Load the incubator so that the sample material is only within the uti-
lized space and make sure there is sufficient air circulation inside the
incubator.