159 / 174
en
12.2020 / Rev. 02
AB66005
Examples
Customer examples and practical
wishes
B
Examples
In manual mode: Temperature drop
■
During the entire weekend (Friday afternoon to Monday morning)
■
Over one day (Sunday)
■
With holiday function (using Wednesday as example)
■
Over the weekend with start delay
In program mode
■
External devices are integrated in program mode, e.g. DAY/
NIGHT simulation.
■
Combination of standstill mode with incubation mode (energy-
saving mode)
■
Incubator is always cleaned on Wednesdays due to the laboratory
routine (SOP) (standstill mode)
B.1
Temperature drop during the entire weekend (Friday afternoon to Monday
morning)
Mon
Tue
Thu
Fri
Sat
Sun
Wed
37
°C
4
Mon
Fig. 219: Diagram ‘Temperature drop during the entire weekend
(Friday afternoon to Monday morning)’
Operating mode:
Manual mode; Temperature drop via real-time calendar
Procedure:
1.
In manual mode, set the temperature drop to 4°C from Friday
afternoon at 2:00 p.m. to Monday morning at 3:00 a.m. and
set the repetition interval to
‘weekly’
2.
Set the target temperature to 37°C
3.
Start manual mode
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...