en
12.2020 / Rev. 02
AB66005
136 / 174
Technical data
Maximum load per standard feed
50 kg
Maximum total load
80 kg
1)
Without door handle and access port (
Ø
42 mm).
2)
With Switchboard option (external socket strip)
3)
With glass door +6 kg.
4)
Lowest settable temperature -5 °C. Reaching a temperature < 0°C
depends on the ambient conditions, however.
All temperature data was measured at an ambient temperature of
22°C and in acc. with DIN 12880:2007-05. The data applies to incu-
bators with standard equipment.
9.3
HettCube 400
Manufacturer
Andreas Hettich GmbH & Co. KG, D-78532 Tuttlingen
Model
HettCube 400
Type
64000
64000-01
Mains voltage (
±
10 %)
220-240 V 1
~
100-120 V 1
~
Mains frequency
50-60 Hz
Connected load
450 VA/850 VA
2)
500 VA/900 VA
2)
Energy consumption at 37°C
0.046 kWh/h
Inner volume
310 l
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Height
Up to 2000 m above sea level
Ambient temperature
16°C to 35°C
Humidity
Maximum relative humidity 75%, non-condensing.
Overvoltage category (IEC
60364-4-443)
II
Degree of contamination
2
Device protection class
I
Noise level
≤
41 dB(A)
Not suitable for use in explosion-endangered areas.
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...