49 / 174
en
12.2020 / Rev. 02
AB66005
Operation
Fig. 55: Program mode
6.8.3
Functional description of program mode
Via program mode, you can combine complex sequences of temper-
atures and durations to form one program.
A program is composed of program segments. A program segment
can also be defined as a standstill segment (incubator in standstill
without regulation).
Each program segment consists of a temperature and a duration.
The duration is calculated via the start time and end time or the end
time is calculated from the start time and duration.
It can be read off from the rotating circular ring in which program
segment the program is currently located. Firstly, the number of the
currently run-through program segment is displayed in the middle of
the circular ring. Secondly, the circular ring itself is subdivided into as
many parts as there are program segments. The number of circular
ring segments shown in black corresponds to the number of the pro-
gram segment that is currently being run.
Example:
The program consists of 4 segments.
−
When the first segment is being run through, the
length of the black filled-out segment of the cir-
cular ring is ¼ of the total circle.
−
When the second segment is being run through,
the length is ½ of the total circle, etc.
−
When the last segment is being run through, the
rotating circular ring is filled out completely in
black, as shown on the screen above.
In the case of cooled incubators, each program segment can addi-
tionally control the control contact (On/Off)
trol contact (option)’ on page 101.
By saving the program, only the temperatures and durations of the
program segments are saved. During the planning, a start date and
start time can be defined. This way, a daily routine or a specification
can be simply programmed.
The following additional functions can be activated by starting a pro-
gram:
■
Program
‘at temp.’, program ‘at start.’
■
After the program runs, specify a hold temperature.
■
Activate the control contact (for cooled incubators).
■
Time-delayed start of program mode.
■
Repeat program.
If the incubator is switched off while manual or pro-
gram mode is running, when the incubator is
switched on, it indicates that a power failure was
registered.
If program repetitions were defined in the start settings, these are
executed
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...