Operation
Class 3.1
Fig. 152: ‘Temperature guard’ menu,
class 3.1
Triggering example, class 3.1
Class 3.2
switched off; for class 3.2, the cooling. As soon as the actual tem-
perature is back within the limits, the incubator will continue to regu-
late. The temperature guards serve as device protection and sample
protection. The temperature limits can limit the standard temperature
range.
In the "Temperature guard" settings, temperatures for classes 3.1
and 3.2 can be input.
For 3.1, up to +70 °C
For 3.2, from -10 °C up to +70 °C
If classes 3.1 and 3.2 are set, they must lie at least 4°C / 8°F apart.
It is only sensible to use a temperature guard of class
3.1 in heating mode with monitoring temperatures
which lie above room temperature.
The temperature specification value can be changed by tapping it.
The setting ex works is +70 °C. This temperature lies outside of the
temperature range of the incubator. Class 3.1 here acts like device
protection, but is not indicated as being active on the main screen. If
the setting is changed, the class 3.1 is indicated as being active on
the main screen.
■
Target tempe37 °C, class 3.1 at +40 °C
Fig. 153: Diagram of ‘Temperature guard’, class 3.1
A Set temperature
B Temperature specification, class 3.1
C Deviation from the target temperature starting from time C
D Class 3.1 active starting from time D
It is only sensible to use a temperature guard of class
3.2 in cooling mode with monitoring temperatures
which lie below room temperature.
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...