139 / 174
en
12.2020 / Rev. 02
AB66005
Technical data
Spatial temperature deviation at
25 °C
±
0.1 K
Recovery time (after door open for
30 s) at temperature setpoint 37
°C
≤
4.5 min
Dimensions
Interior dimensions
W x D x H
535 x 690 x 850 mm
Exterior dimensions
W x D
1)
x H
710 x 825
1
x 1425 mm
Weight
approx. 128 kg
3)
Maximum load per standard feed
50 kg
Maximum total load
100 kg
1)
Without door handle and access port (
Ø
42 mm).
2)
With Switchboard option (external socket strip)
3)
With glass door +10 kg.
4)
Lowest settable temperature -5 °C. Reaching a temperature < 0°C
depends on the ambient conditions, however.
All temperature data was measured at an ambient temperature of
22°C and in acc. with DIN 12880:2007-05. The data applies to incu-
bators with standard equipment.
9.5
HettCube 600
Manufacturer
Andreas Hettich GmbH & Co. KG, D-78532 Tuttlingen
Model
HettCube 600
Type
66000
66000-01
Mains voltage (
±
10 %)
220-240 V 1
~
100-120 V 1
~
Mains frequency
50-60 Hz
Connected load
450 VA/850 VA
2)
500 VA/900 VA
2)
Energy consumption at 37°C
0.056 kWh/h
Inner volume
520 l
Ambient conditions (EN / IEC 61010-1)
Set-up site
Indoors only
Height
Up to 2000 m above sea level
Содержание HettCube 200
Страница 151: ...151 174 en 12 2020 Rev 02 AB66005 Index W Warning symbols 7 Warnings and error messages 126 ...
Страница 152: ...en 12 2020 Rev 02 AB66005 152 174 Index ...
Страница 153: ...13 Appendix ...
Страница 154: ......
Страница 166: ...en 12 2020 Rev 02 AB66005 166 174 Examples ...
Страница 170: ...en 12 2020 Rev 02 AB66005 170 174 Accessory overview ...
Страница 171: ...171 174 en 12 2020 Rev 02 AB66005 Declaration of conformity D Declaration of conformity ...