Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
4.
Remove the optional left and right safety
arch. Keep the screws.
5.
Dismount the motor cover.
6.
Open the cable tray cover on the side of the
load cell to replace.
7.
Remove the tie-wraps on the load cell cable.
Remember the position of the tie-wraps.
8.
Dismount 26x floor anchoring screws. Keep
the screws and washers.
9.
Disconnect the appropriate load cell cable
from the amplifier.
10.
Dismount both, left and right, handrail from
the handrail supports, set screws and
washers aside, then lay the handrails
temporarily on the treadmill.