gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 61
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
18.
Ensure that all criterions are passed and
press “Validate”. Otherwise repeat the steps
of a given position that show the largest
errors or repeat all steps from the beginning.
19.
In the gaitway-3D[Service] software, select
“Fore-aft force sensors calibration”.
20.
Check that nothing touches the treadmill.
21.
Fix the calibration gauge in position 1: gauge
axis under the right side of the rear roller.
Use two F-clamps of 80mm. Do not tighten
excessively the clamps. The gauge cable is
oriented to the floor. There can be no
tension on the gauge cable. Check that the
clamps do not touch any other parts than the
calibration gauge and the treadmill lateral
beam under the rear roller.