gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 15
For the calibration of the fore-afterward ground reaction force channels, a calibration
gauge is fixed at different positions behind the treadmill. A rope is used to pull on the
calibration gauge. The force is measured simultaneously by the instrumentation
amplifier and the calibration amplifier.
For the calibration of the lateral ground reaction force channels, the calibration gauge
is placed at different positions along the side of the treadmill. A rope is used to pull on
the calibration gauge. The force is measured simultaneously by the instrumentation
amplifier and the calibration amplifier.
The software then computes the 6x8 calibration matrix for the three ground reaction
force direction and the three moment direction from the 8 measurement channels.
Figure 12. Calibration network