![background image](http://html.mh-extra.com/html/h-p-cosmos/arsalis-gaitway-3d/arsalis-gaitway-3d_service-manual_574521075.webp)
gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 71
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
8.
In the gaitway-3D[Service] software, select
“Checkout: Export Flash & data” and follow
the instructions to save a zip package with
the calibration and checkout data and
results. Save it on a USB stick and send a
copy to
The gaitway-3D[Service] software can now
be closed.
9.
Start gaitway-3D software and report
software version.
Version:
10.
Enter "Measure" function to connect to the
gaitway-3D treadmill and amplifier.
11.
Enter in Hardware control screen and report
(copy-paste) the Amplifier information and
the treadmill info
Amplifier info
Treadmill Info
12.
Enter in Operator screen and select/create
one operator record.
13.
Enter in Patient screen and select/create
one test patient record.