gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 62
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
22.
Follow the instructions of the gaitway-
3D[Service] software.
To pull on the calibration gauge, use a rope
and a lever. The rope is attached to the lever
with a well tightened knot. The distance from
the knot to the lever extremity should be
equal to the height from the floor to the
eyebolt of the gauge to pull horizontally.
Place the extremity of the lever on the
ground against the feet to pull on the rope.
Take care to pull in the axis of the treadmill.
When a zero load is requested let the rope
loose.