![background image](http://html.mh-extra.com/html/h-p-cosmos/arsalis-gaitway-3d/arsalis-gaitway-3d_service-manual_574521059.webp)
gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 55
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
78.
Connect the Ethernet cable from the
gaitway-3D force amplifier to one router LAN
port.
Ethernet cable
79.
Connect the RS-232 cable between the
treadmill (user terminal) and the computer.
If the optional USB to RS-232 converter is
used, install the driver on the computer.
RS-232 cable
Optional: USB to RS232 adapter,
Digitus DA-70156, cos12796-01
80.
Connect the treadmill mains power plug.
81.
Switch on the treadmill.
82.
For MCU5 user terminals:
From the treadmill user terminal, enter the
user options by pressing the following
buttons simultaneously for 3 seconds:
“SPEED – (<F3>)”
“SPEED + (<F4>)”
“STOP (<F8>)”
Select USER-OPTION: 20
Set the value to 20 (default is 1)
Press ENTER to validate
Press STOP to quit user-options
This sets the treadmill COM 1 to baud rate of
115200 baud
instead
of
the
default
9600 baud