gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 103
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
11.
Measure the impedance of the load cell 4
power supply on the load cell cable (pin1) to
- the load cell ground (pin6)
- the cable connector backshell
R
GND4
=
333.3ohm
±5ohm
R
backshell4
>
1Mohm
R
GND4
=
R
backshell4
>
12.
Reconnect the 4 load cells and tighten the
connector screws with a screwdriver.
13.
Reconnect the power cable.
14.
If all the steps are passed successfully, the
fuse can be replaced as follow:
15.
Unscrew the fuse holder cap.
16.
Check that the old fuse is blown out.
17.
Replace the fuse by a new fuse 800 mA, 250
V, time delay, 5mm x 20mm such as
SCHURTER 0034.3116.
18.
Screw the fuse holder cap with the new fuse.
19.
Insert the main plug of the treadmill and start
the treadmill.
20.
Confirm that the amplifier status led is on.
21.
Shut down the treadmill.
22.
Reconnect the accessories disconnected at
step
23.
Start the treadmill.
24.
Confirm that the amplifier status led is on.
END OF PROCEDURE