![background image](http://html.mh-extra.com/html/h-p-cosmos/arsalis-gaitway-3d/arsalis-gaitway-3d_service-manual_574521025.webp)
gaitway-3D Service manual
Ref.:TM-MAN-0005-ARS
Iss.: A
Rev.: 9
Date: 07/12/2020
Page : 21
Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
8.
Place the sensor cases on the base plate.
Screw each of them with 4 screws and
washers. Attach a tie-holder on the rear-
internal side each load-cell.
Torque =
1.8Nm
4x Capteur-Mini3D-Case
16x ISO7380 M4x10,
16x DIN6798A M4,
4x tie-holder
9.
Tighten the cable glands on the cables.
10.
Screw the Flat-Support on the sensors with
4x DIN7991-M12x25.
Torque =
50Nm
4x Flat-Support
4x DIN7991-M12x25
Note: Ensure that each load cells is
tightened at its nominal torque on the
first mounting side before tightening the
second mounting side, to avoid
damaging the insert on load cell support
part.