Test Step
Measurements/Test Results
Seq.
No.
Description
Required/No
minal (incl.
tolerance)
Measured/
Calculated
Remarks/Calculations
Pass/
No
pass
23.
Position 2 is with gauge axis at ¼ of belt
width from the right side.
Position 3 is in the center of the belt.
Position 4 is with gauge axis at ¼ of belt
width from the left side.
Position 5 is with gauge axis under the left
side of the rear roller.
24.
Ensure that all criterions are passed and
press “Validate”. Otherwise repeat the steps
of a given position that show the largest
errors or repeat all steps from the beginning.
25.
In the gaitway-3D[Service] software, select
“Lateral force sensors calibration”.
26.
Check that nothing touches the treadmill.