ORALIX AC SYSTEM
TECHNICAL DATA
4.3.
CO
//,0$725
Collimator type
metal cone, lead shielded, bayonet fitting
Distance from Focal Spot to distal cone end
20 cm
Irradiated field at distal cone end
6 cm diameter (circular, standard)
3 x 5 cm
(rectangular, optional)
4.4.
TIMER
Versions
·
Densomat, object programmed with +/- manual time
correction, with possibility to control 2 Tubeheads by
optional extension hardware.
·
Secondent, with manual time setting by time selec-
tor.
Selectable exposure times
among 19 settings, according to R10 scale:
0.03, 0.05, 0.06, 0.08, 0.10, 0.12, 0.16, 0.20, 0.26, 0.32,
0.40, 0.50, 0.64, 0.80, 1.00, 1.26, 1.60, 2.00, 2.50.
Actual exposure time
·
For 50 Hz is the number of X-ray pulses correspond-
ing to the selected time.
·
For 60 Hz is the number of X-ray pulses with mini-
mum deviation respect to the selected exposure
time.
·
the exposure time is measured between 25% of the
level of the anode current waveform.
The accuracy is +/-(10% +1 ms)
“X-ray ON” indication
visual (LED) and audible (buzzer)
Preheating time
120 msec (default) 6 pulses @ 50 Hz, 7 pulses @ 60 Hz
Compensation of mains fluctuations
by automatic correction of exposure time
4.5.
WEIGHTS
Wall System
25 kg
Mobile System
60 kg
Tubehead Oralix AC, 230V
6.3 kg, without collimator
Folding Arm
7.5 kg
Extension Arm
45 cm
3.5 kg
90 cm
5.0 kg
Wall Support
1.5 kg
Densomat Timer
2.0 kg
Secondent Timer
2.0 kg
For the dimensions of the system see AZ-3 and AZ-4