INSTALLATION
B-6
Gendex Dental Systems
ORALIX AC SYSTEM
2.2.
INSTALLATION TO THE WALL
Place the Wall Support and the Timer at the desired location on the wall and mark the holes in the wall,
using the Wall Plate as a template
... and the template on page BZ-2 for the Timer.
WARNING
Do not use the Timer as
a template ! Electronic
components could be
irreparably damaged !
Drill the four pilot holes:
- for the Wall Support (with a 9 mm drill bit)
- for the Timer (with a 8 mm drill bit).
Position the Timer over the four pilot holes, hav-
ing routed the line cable through the Timer knock
out and having routed additional cable from the
Timer to the Wall Support.
1
additional cable
Timer-Wall Support
line cable
(mains)
3
2
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER