Gendex Dental Systems
B-11
ORALIX AC SYSTEM
INSTALLATION
h
k
b
2.4.
MOUNTING OF FOLDING ARM
Unpack the Folding Arm taking care not to re-
move its bindings before the Tubehead is installed.
WARNING
The springs in the Folding Arm are powerful and
may cause injury to installer (as well as possible
damage to the arm) if not handled properly. Do
not remove binding of Folding Arm until indi-
cated in the instructions below.
Slide the chromium ring (h) over support pin (b),
fitting the ring hole into locking pin (k).
If necessary set sector (i) (not supplied, only
available as spare part) into the track provided of
Extension Arm, to limit horizontal rotation.
REMARK
Set in position A, to limit right hand rotation.
Set in position B, to limit left hand rotation.
optional sector (i)
code 4522 121 93181
1
2
3
Danger !
Do not cut the
holding safety
bindings