Gendex Dental Systems
B-21
ORALIX AC SYSTEM
INSTALLATION
X10
X11
4. OPTIONS
4.1.
SECOND TUBEHEAD INSTALLATION
When the Oralix AC System is equipped with the Timer DENSOMAT, a second Tubehead can be installed and
driven by the same DENSOMAT Timer console.
Mechanical installation. The me-
chanical installation of a second
Tubehead is equal to the one al-
ready described for the single
Tubehead; therefore, for instruc-
tions refer to Chapters 1 e 2 of this
section B.
An optional plug-in card is provided-on-
request in order to enable the driving of a
second Tubehead. To install the card in
the DENSOMAT Timer follow the instruc-
tions below:
Carefully plug-in the option card on
X10 and X11 connectors in the
Densomat printed circuit board.
Set the dipswitch S1-2 from the OFF
position (single tubehead) to the ON po-
sition (double tubehead) to enable the
second tubehead.
Second Tubehead electrical con-
nection. The mains electrical con-
nections have to be carried out on
X2 terminal block through the fol-
lowing pins:
- X2:1 (live);
- X2:2 (neutral).
The earth wire (y/g) must be con-
nected to X3 terminal block to the following pin:
- X3:2 (ground).
4.2.
EXTERNAL LIGHTS
The Oralix AC system can drive external lights indicating the ready for exposure and the X-ray emission in
progress: an optional board (code 4519 101 02601) is available for external lights connection.
See principle diagrams Z-3 and Z-4 for the connections.
2