ORALIX AC SYSTEM
CORRECTIVE MAINTENANCE
F-1
Gendex Dental Systems
SECTION F
CORRECTIVE MAINTENANCE
1.
INTRODUCTION ................................................................................................ F-1
2.
EQUIPMENT REQUIRED .................................................................................. F-1
3.
ADJUSTMENTS ................................................................................................ F-2
3.1.
FOLDING ARM FRICTION ............................................................................................................ F-2
3.2.
FOLDING ARM SPRING TENSION ............................................................................................... F-3
3.3.
TUBEHEAD FRICTION .................................................................................................................. F-3
4.
REPLACEMENT PROCEDURES ..................................................................... F-4
4.1.
REPLACING THE WHEELS OF THE MOBILE STAND ................................................................. F-4
5.
ADJUSTMENT OF THE MAINS VOLTAGE MEASURING CIRCUIT ................ F-5
5.1.
ADJUSTMENT PROCEDURE ON THE TIMER DENSOMAT ......................................................... F-5
5.2.
ADJUSTMENT PROCEDURE ON THE TIMER SECONDENT ....................................................... F-5
6.
TROUBLESHOOTING ....................................................................................... F-6
6.1.
SYSTEM EQUIPPED WITH TIMER DENSOMAT .......................................................................... F-6
6.2.
SYSTEM EQUIPPED WITH TIMER SECONDENT ........................................................................ F-7
3523(5',6326$/2)(/(&7521,&(48,30(17)
1. INTRODUCTION
The Corrective Maintenance provides information about adjustment and replacement procedures concerning
the Oralix AC system.
WARNING
Always disconnect the equipment from the mains supply before starting any maintenance action:
the mains voltage may be present on parts of the system (such as electronic board, Tubehead,
Folding Arm cable) even in the stand-by condition.
In case of replacement/maintenance of Tubehead, Folding Arm and Extension Arm it is neces-
sary to park the Folding Arm in a safe position by guiding the arm slowly to the horizontal position
opposing by hand to the springs action directed upwards (see figure in pag.B14).
If the Folding Arm is released without Tubehead and without opposing by hand to the springs
action, it is possible to damage the springs themselves.
2. EQUIPMENT REQUIRED
Standard service toolkit;
Special drift code 4519 190 00871;
Loctite 290 code 1322 511 00702;
Special 30 mm open wrench code 4519 120 93161 (ONLY FOR MOBILE UNITS);
Special 30 x 17 mm open wrench code 4519 120 92141 (ONLY FOR MOBILE UNITS).