INSTALLATION
B-10
Gendex Dental Systems
ORALIX AC SYSTEM
grease here
(c)
Lightly grease support tube of Extension Arm.
Feed guide string and insert Extension Arm into
opening at top of Wall Support.
By using the nut and lock washer supplied, con-
nect one side of the protective ground wire (b) to
the ground point of the Extension Arm, then con-
nect the other side to the earth screw of Wall
Support.
Loosen two friction screws (c) of Extension Arm.
Remove round snap-out covers (d) at both ends
of Extension Arm (see figure in the previous page).
ground point of the Ext. arm
b
earth screw of the Wall Support
4
5
6
7
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER