INSTALLATION
B-12
Gendex Dental Systems
ORALIX AC SYSTEM
Tie the guide cord of the Extension Arm to the Folding Arm cable.
Feed the Folding Arm cable through the Exten-
sion Arm and through the Wall Support, with the
help of the guide cord.
Lightly grease the support pin
By moving vertically, carefully insert the support
pin of the Folding Arm into the Extension Arm;
secure finally the screws (c) of the Extension Arm.
Connect the wires of the Folding Arm cable to the
Wall Support terminal strip, as described in BZ-
4, System electrical connections.
4
5
6
(c)
7