INSTALLATION
B-2
Gendex Dental Systems
ORALIX AC SYSTEM
1. WALL MOUNTED SYSTEMS
1.1.
ROOM PREPARATION
The Wall Support shall be mounted either on a side-wall or on back-wall respect to the patient, outside of the
dentist operational area.
The maximum distance from the center of the patient head is 170 cm (135 cm with the short Extension Arm)
in order to ensure the needed reach for all the possible tecniques; the recommended height from the floor is
133 cm in order to guarantee the easy positioning either with patient sitting or lying down.
NOTE
The static load applied by the system on each attachment point is 72 Kg (159lb).
Considerating a safety factor of four, it is specified that each attachment point is able to
support pull and thrust load of 288 Kg (636lb).
REMARK
It is task of the installation engineer to choose the proper fixation hardware in relation to
the type of the mounting wall.
WARNING
Insufficient wall or hardware strength may cause the wall mount to pull out of the wall, and
the arm and Tubehead to fall, causing serious injury to the patient or operator.
1.2.
WALL MOUNTING METHODS
Wall mountig by means of expansion screws.
The Wall Support is delivered with a set of expansion screws
(M6x60, diameter 11 mm) only suitable for use with concrete
walls type R 300 and over (resistance to compressive load >300
Kg/cm
2
).
In case that such a wall is available, drill the mounting holes
(diameter 12 mm) in the desired location of the wall, making
use of the template BZ-1.