TECHNICAL DATA
ORALIX AC SYSTEM
Gendex Dental Systems
A-6
4.
TECHNICAL DATA
4.1.
MAINS SUPPLY
Mains voltage
230 VAC +/-10%, single phase (polarized) or double phase
it is reccomended to use a dedicated magneto/thermal circuit breaker (10
A max) and an earth leakage circuit breaker (leakage current of 30 mA)
Current absorption
$0D[LPXP$6WDQGE\
Mains fuse
F6.3A / 250V, fast blow 2
nd
fuse for double phase, with by-pass jumper
Frequency
50 Hz / 60 Hz ± 2 Hz
Line resistance
0.8 Ohm
4.2.
TUBEHEAD
Peak tube potential (see section D-3.2.)
60.0 kVp +/- 8%
(Nominal mains Voltage -10%)
65.0 kVp +/- 8%
(Nominal mains Voltage)
68.5 kVp +/- 8%
(Nominal mains 10%)
Tube current (see section D-3.3.)
6.
mA +/- 10%
Nominal mains Voltage -10%)
mA +/- 10%
Nominal mains Voltage)
.
mA +/- 10%
Nominal mains V10%)
Radiation output
5 mGy/sec +/-30% at 20 cm from the focus
Tube rating chart (only 1 fixed working point)
65 kVp,
.
5 mA, 2.5 s
X-ray tube
800-000
G
, grid controlled, single focus
DQG
X-ray Tubehead
Oil immersed, shielded
Minimum filtration permanently in the beam
2.0 mm Al equiv. @ 65 kVp (IEC 522/1976)
Beam quality half value layer (H.V.L.)
better than 1.7 mm Al @ 65 kVp
Focal spot
0.7 mm (IEC 336/1982)
Duty cycle
1
Minimum wait time
3 sec
Radiation leakage
less than 4
microGy/h (5 mR/h) @ 1 meter
Leakage Tecnique factors
0.1 mA @ 65 kVp
Anode angle
19.5°
+/- 0.5
respect to the
UHIHUHQFH
axis
p
XOVH
7DUJHWPDWHULDO
7XQJVWHQ