Gendex Dental Systems
A-5
ORALIX AC SYSTEM
TECHNICAL DATA
When the exposure button is depressed, the X-ray ON indica-
tor lights and an audible signal sounds. The exposure is ter-
minated when the exposure button is released or when the
programmed exposure time is completed.
The Secondent automatically compensates for line voltage
fluctuations, consistently providing films with optimal density.
(The voltage compensation can be inhibited by means of a
dedicated jumper).
The Secondent offers the possibility to select different Tubehead
versions (220-230-240 VAC)
The Timer provides 3 LED indicators for: “ready for exposure”
(green), “X-ray emission” (yellow) and “alarm” condition (red).
By optional extension hardware it is possible to connect
external lamps for “ready for exposure” and “X-ray emission”
indications.
3.3.
TUBEHEAD ORALIX AC
The X-ray Tubehead Oralix AC (230 V) consists of a high voltage and filament transformer, and 3-elements,
grid-control X-ray tube, in a lead shielded, oil filled metal case, covered by a plastic housing. A collimator/
positioning device with quick connect bayonet fitting is supplied with the Tubehead.
3.4.
FOLDING ARM
The Folding Arm is a 2-section spring counter-balanced dental X-ray arm. The end of the Folding Arm has a
3-conductor cable which interconnects with the Densomat/Secondent Timer via the Wall Support, when the
system is assembled.
3.5.
EXTENSION ARM
An Extension Arm available in 45-90 cm is provided with the system. Electrical wiring and mechanical
interconnection between the Folding Arm and the Wall Support is made via the Extension Arm.
3.6.
WALL SUPPORT
The Wall Support incorporates a mechanical fitting into which the support tube of the Extension Arm is installed.
A terminal strip is provided for wiring interconnection between the Densomat/Secondent Timer and the Folding
Arm and Tubehead.