SETTING TO WORK
C-5
Gendex Dental Systems
ORALIX AC SYSTEM
5.2.2. Tubehead Voltage Select routine
This routine is provided for installation of the older types of 220 V and 240 V tubeheads. The Timer AC is only
provided with 230 V tubeheads and so the 230 V default value should not be changed.
After having completed the “Test mode routine” (by pressing the “P” key) the “Tubehead Voltage select routine” is
entered and “SEL” will be displayed for a second.
It is possible to change the nominal mains voltage for the type of tubehead being used. Select the tubehead via
the TUBEHEAD 1 and TUBEHEAD 2 keys and change the voltage by pressing the S+ and S– keys.
Pressing the “P” key exits and “VAC” is displayed.
5.2.3. Measurement of mains voltage (DVM routine)
After having completed the “Tubehead Voltage Select routine” (by pressing the “P” key) the “DVM routine”
(Digital Voltmeter) is initiated and “UAC” (VAC) will be displayed for a second.
The “DVM routine” is a test function used to check the mains voltage value measured by the Timer. This value
is shown on the display.
During “DVM routine”, every 1 second the display is updated showing actual value of mains voltage with an
accuracy of ± 1V.
Pressing “P” key exits “DVM routine” and the display shows “cnt”.
5.2.4. Exposure Counter routine
After having completed the “DVM” routine (by pressing the “P” key) the “Exposure Counter routine” is entered
and “cnt” will be displayed for a second.
It is possible to view the number of exposures made for the selected tubehead. This number is displayed in
multiples of ten; e.g. if 015 is shown on the display than between 150 and 160 exposures have been made.
Select the tubehead via the TUBEHEAD 1 and TUBEHEAD 2 keys.
Pressing “P” key exits and the display shows “END”.
Switch off the System.
“P”
key
DENSOMAT TIMER
- Front view -
“Vx”
key
“bitewing”
key
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER