Gendex Dental Systems
C-2
ORALIX AC SYSTEM
SETTING TO WORK
3. CONTROLS, INDICATORS AND JUMPERS SETUP
For a survay of the controls and indicators of the Timers see CZ-1 and CZ-2.
3.1.
JUMPERS AND DIP-SWITCHES SETUP
The table below describes the jumpers and dip-switches setup on Densomat/Secondent Timer.
DENSOMAT
SECONDENT
Factory
Custom setting
Factory
Custom setting
default setting
default setting
W1
Not used
Not used
Not used
Not used
W2
1-2 (B.U.T. enabled)
2 - 3 (B.U.T. test)
1 - 2
Not used
W3
Unplugged
Not used
Opened
Not used
W4
Plugged
Not used
Closed
Not used
W5
Closed
Not used
Closed
Not used
W6
Opened
Not used
Opened
Not used
W7
Opened
Not used
Opened
Not used
W8
1-2 (X-ray enabled)
2-3 (X-ray inhibited)
1-2 (X-ray enabled)
2-3 (X-ray inhibited)
W9
Opened
Not used
Opened
Not used
W10
Closed
Not used
Closed
Not used
W11
Closed (F3 shunted)
Opened (F3 working)
Closed (F3 shunted)
Opened (F3 working)
S1 DENSOMAT
SECONDENT
1
not used
see table
“Tubehead version
2
OFF=
single Tubehead (*)
selection”
ON =
double Tubehead
3
ON= compensation
OFF= compensation
(**) inhibited (*)
inhibited
OFF = compensation
ON =compensation
enabled
enabled (*)
4
OFF= X-ray inhibited
OFF= X-ray inhibited
ON = X-ray enabled (*)
ON = X-ray enabled (*)
8
not used
not used
(*) = DEFAULT (**) = Manual mode only
(Exposure time selected manually by pressing S+/S- pushbuttons)
Preheating time selection at 50Hz
(applicable for Densomat & Secondent)
TIME
S1
(ms)
5
6
7
80
OFF
OFF
OFF
100
OFF
OFF
ON
120 (*)
OFF
ON
OFF
140
OFF
ON
ON
160
ON
OFF
OFF
180
ON
OFF
ON
200
ON
ON
OFF
220
ON
ON
ON
J
U
M
P
E
R
Tubehead version selection
(Secondent only )
(*)= DEFAULT
Tube version 230V (*)
240V 220V spare
S1 - 1
ON
OFF
ON
OFF
S1 - 2
OFF
ON
ON
OFF